62 Setting PowerLine Controller Operating Parameters
Active PowerLine Devices
The System Configure screen indicates which PowerLine devices are active on the
IEEE-488 bus. This screen displays device boxes that include the device names and IEEE
addresses of all PowerLine devices that the system recognizes.
3. Verify that the PowerLine devices installed in your system appear on the System
Configure screen.
4. If a connected PowerLine device is not listed on the System Configure screen, press
the Rescan System screen key. The controller rescans the IEEE-488 bus to locate
active devices.
The controller must rescan the system to recognize PowerLine devices and to make
them active if:
You powered up a PowerLine device after you powered up the controller
You powered up the controller before a PowerLine device finished its powerup
calibration sequence
5. If rescanning the system does not list all PowerLine installed devices, check your
IEEE-488 connections. See the Waters 600E Multisolvent Delivery System
Installation and Maintenance Guide, Section 3.3, Making IEEE-488 Interface
Connections, for the PowerLine device connection procedure.
Note: If you do not set a PowerLine device IEEE-488 address correctly, the
controller will not recognize the device. See the Waters 600E Multisolvent Delivery
System Installation and Maintenance Guide, Section 3.3, Making IEEE-488
Interface Connections, for information on setting addresses and powerup
sequences.
6. Press the Return screen key to exit the System Configure screen and return to the
Pump Setup screen.
7. If you are setting up your system for the first time, or if you wish to change one or
more pump parameters, proceed to Section 4.3, Setting Up the 600E Pump.