140
7. Insert the fuse holder into the receptacle, and gently push until it locks
into position.
Tip: A clicking sound indicates the fuse holder is locked into position.
8. Connect the power cord to the rear panel power entry module.
Cleaning the instrument’s exterior (all models)
Clean surfaces of the column heater, column manager, auxiliary column
manager, or column heater/cooler with only a soft, lint-free paper or cloth
dampened with water.
Observe these requirements when cleaning device surfaces:
Always ensure the electrical power to the device is interrupted.
Always use eye and hand protection during the cleaning process.
Apply the water to a clean cloth only, and then wipe the device.
Never spray or apply the water directly onto any device surface.
Warning: To prevent burn injuries, adjust the temperature set-point of
the column manager to a safe temperature (10 to 30 °C) and allow the
unit to attain this set-point temperature before handling components
within the column compartment. When the column manager is set at a
high temperature, all interior surfaces and components are hot.