Preparing the column compartments 75
Sample
manager
injector valve,
port 6
CM-A inlet
valve, center
port (I)
19
b
430002599, tube
assembly,
.004-inch ID ×
19.0-inch long
(reorder part
number
700005479)
430002641
(reorder part
number
700005484)
CM-A outlet
valve, center
port
Detector inlet 14.5 430002681,
tubing,
.004-inch ID
(reorder part
number
700005477)
430002682
(reorder part
number
700005482)
CM-A outlet
valve, center
port
Detector inlet 19 430002683,
tubing,
.004-inch ID
(reorder part
number
700005478)
430002684
(reorder part
number
700005483)
CM-A outlet
valve, center
port
Detector inlet
22.5
b
430002657,
tubing,
.004-inch ID
(reorder part
number
700005480)
430002658
(reorder part
number
700005485)
a. Be sure to connect the 430002727 waste tubing to port W (at the 3 o’clock position) of the inlet
valve. Do not connect to the center port.
b. Depending on your configuration, it may be necessary to order extra lengths of this tubing.
Four and six column configuration plumbing connection details: (Continued)
From To
Length
(inches)
Part number
Stainless steel
MP35N
(H-Class Bio)