Item Specification
Maximum sample plate
height (includes vials,
caps, and cap mats)
53 mm
Sample compartment
temperature range
Between 4 and 40 °C, in increments of 0.1 °C, with a tolerance range
of between -2 and +4 °C
• At a setpoint of 4 °C with ambient temperature <23 °C and
humidity <80%, maintains a sample temperature of 2 to 8 °C.
• At ambient temperatures >23 °C and/or humidity >80%, the
sample manager and sample organizer can maintain an average
sample temperature of 18 °C below ambient, ±3.0 °C.
Temperature accuracy No more than ±1.0 °C in temperature between a traceable external
temperature measurement device and an instrument temperature
measurement device.
Temperature stability ±1.0 °C (at the sensor, with the sample compartment door closed)
Pneumatic system
operating pressure
range
414 to 758 kPa (4 to 8 bar, 60 to 110 psi)
Optional external
pneumatic source
pressure range
517 to 689 kPa (5 to 7 bar, 75 to 100 psi)
August 8, 2016, 715005049 Rev. C