EasyManua.ls Logo

Waters ACQUITY - Connecting and Disconnecting the External Source Exhaust Valve Assembly (KAB)

Waters ACQUITY
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Warning: Peroxide contaminants in THF can spontaneously and destructively explode
when you partially or completely evaporate the THF.
Warning: Hexane is a neurotoxin, and THF can irritate eyes, skin, and mucous
membranes and cause harmful neurological effects. If you use either or both of these
volatile solvents, locate your system inside a fume hood or walk-in chamber to minimize
exposure to harmful solvent vapors.
THF, hexane, ethyl acetate, and acetone can be used as the mobile-phase in ACQUITY
UPLC H-Class systems. However, as with many nonaqueous solvents, they can shorten
system and instrument life compared with equipment running typical reversed-phase solvents.
If you routinely use THF, hexane, ethyl acetate, or acetone, Waters recommends you install
the Hexane/THF Compatibility Kit.
When using unstabilized THF, ensure that your solvent is freshly prepared. Previously opened
bottles contain peroxide contaminants, which cause baseline drift.
Chloroform, methylene chloride, halogenated solvents, and toluene are generally not
recommended for use in ACQUITY UPLC H-Class systems. Nevertheless, you can use these
solvents in weak dilutions (<5%) as additives, sample diluents, or modifiers.
Contact your Waters sales representative or local technical support organization to determine
whether a specific method is suitable to use with the system instruments and components.
When using THF or hexane, install stainless steel tubing, and minimize the use of PEEK
components.
Aqueous solvents must not remain in a shut-down system because they serve as a substrate
for microbial colonies. Microbes can clog system filters and capillary lines. To prevent their
proliferation, add a small amount (~10%) of an organic solvent such as acetonitrile or
methanol.
Methanesulfonic acid is not recommended for use in ACQUITY UPLC H-Class systems.
B.12
Quaternary solvent manager recommendations
The seal wash system must never run dry, particularly during separations that use a polar
mobile phase.
Isopropyl alcohol or mixtures of methanol and water, like 20% methanol/water, are effective
seal wash solvents for THF solvent mixtures.
For reversed-phase applications, use aqueous seal wash solutions with a weak organic
component (for example 1:9 methanol/water).
Do not use 100% organic seal wash solutions.
August 8, 2016, 715005049 Rev. C
Page 91

Table of Contents

Other manuals for Waters ACQUITY

Related product manuals