EasyManua.ls Logo

Waters ACQUITY - Sample Manager - Flow through Needle Recommendations; Common Solvent Properties

Waters ACQUITY
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B.13 Sample manager - flow through needle
recommendations
Do not use concentrations of THF or hexane greater than 10% as purge solvent.
Typical organic sample diluents such as dimethylsolfoxide (DMSO) and dimethylformamide
(DMF) are supported.
Notice:
Do not leave buffers stored in the system.
Flush all flow paths, including the needle wash, with plenty of non-buffered solvent
before shutting down the system.
For extended shutdown periods (longer than 24 hours), use 10% to 20% methanol
in water.
When using a buffered wash solvent, prime it for a minimum of 30 sec.
Use of buffers can cause salt build-up on the needle and wash port, which can
require periodic cleaning.
B.14 Common solvent properties
The following table lists the properties for some common chromatographic solvents.
Table B–1: Properties of common solvents
Solvent Vapor pressure mm Hg (Torr) Boiling point
(°C)
Flash point
(°C)
Acetone 184.5 at 20 °C 56.29 -20
Acetonitrile 88.8 at 25 °C 81.6 6
n-butyl acetate 7.8 at 20 °C 126.11 22
n-butyl alcohol 4.4 at 20 °C 117.5 37
n-butyl chloride 80.1 at 20 °C 78.44 -9
Chlorobenzene 8.8 at 20 °C 131.69 28
Chloroform 158.4 at 20 °C 61.15
Cyclohexane 77.5 at 20 °C 80.72 -20
Cyclopentane 400 at 20 °C 49.26 -7
o-Dichlorobenzene 1.2 at 20 °C 180.48 66
Dichloromethane 350 at 20 °C 39.75
Dimethyl acetamide 1.3 at 25 °C 166.1 70
August 8, 2016, 715005049 Rev. C
Page 92

Table of Contents

Other manuals for Waters ACQUITY

Related product manuals