2.1.1 Follow these general recommendations when performing a
UPLC analysis
Select appropriate solutions
Use high-quality solvents, buffers, and additives (HPLC or MS grade).
Keep concentrated stock solutions, to use when preparing working solutions.
Start gradients that include an organic component (0.1%, for example) to provide more
consistent and predictable gradient formation than when you start with 0% organic.
Ensure that gradients include an organic component (0.1%, for example) to start, to provide
more consistent and predictable.
Set up the system properly
When installing or removing a column, always hold the active preheater’s reusable
compression fitting in place. Rotate the column or optional in-line filter to install or remove it.
Always use solvent filters on tubing lines in solvent bottles.
Use the Load Ahead option when you desire a shorter cycle time.
Prime properly
Prime solvent lines during system start-up.
Keep the seal wash line and all solvent lines primed.
Manage waste properly
Do not block the degasser vent line; trim the tubing, if necessary.
Do not submerge the waste or degasser vent lines in liquid. (See the solvent manager
overview and maintenance guide for details on how to route the tubing.)
Monitor the waste level, to ensure that it is never too high.
Use care when using buffers
Do not use buffers in the wash solvent line.
Do not top off buffers, which can promote microbial growth.
Filter buffers with a 0.2-µm filter membrane.
Do not leave buffers stored in the system.
August 8, 2016, 715005049 Rev. C
Page 27