EasyManuals Logo

Watts AERCO Benchmark BMK5000 User Manual

Watts AERCO Benchmark BMK5000
213 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
Benchmark with Edge [I]: Operation-Maintenance Manual
SECTION 4 INITIAL START-UP
OMM-0145_D GF-218 11/2/2020 Technical Support (800) 526-0288 Mon-Fri, 8 am - 5 pm EST Page 65 of 213
PROPANE Combustion Calibration Instructions
16. With the valve position still at 100%, insert the combustion analyzer probe into the exhaust
manifold probe opening (see Figure 4-2a 4-2c in Section 4.2.3) and allow enough time
for the combustion analyzer reading to stabilize.
17. Compare the combustion analyzer’s oxygen (O
2
) reading to the O
2
value in the Reading
column (Figure 4-9). If they differ, go to the Main Menu Calibration Input/Output
Oxygen Sensor screen and adjust the O2 Offset parameter, up to ±3%, to make the on-
board O
2
sensor match the value from the combustion analyzer. If your combustion
analyzer is correctly calibrated, and the on-board O
2
sensor cannot be made to match the
analyzer, the sensor may be defective and need to be replaced.
18. Compare the O
2
value in the Target and Reading columns. If they don’t match, adjust the
Blower Voltage until the O
2
value in both columns match; use either the + or controls, or
press on the field and type the value directly.
19. If adjusting the blower voltage is not sufficient to get the O
2
Reading column to match the
Target column, then repeat Step 15 to adjust the gas pressure up or down within the range
shown in the table, then repeat Step 18. Continue repeating Steps 15 and 18 until the gas
pressure is within the range in Table 4-4 and the O
2
Reading column to matches the
Target column.
20. Enter the downstream manometer’s gas pressure reading in the Downstream Gas
Pressure field. Note, this field appears only when Valve Position % = 100%.
21. Compare the measured nitrogen oxide (NOx) and carbon monoxide (CO) readings to the
Target values in the Table 4-5 (shown as a reference only). If you are not in a “NOx-
limited” area and/or do not have a NOx measurement in your analyzer, set the O
2
to the
value in the Oxygen (O
2
) % column in the table below.
TABLE 4-5: PROPANE Calibration Readings at 100% Valve Position
Model
Oxygen (O
2
) %
Nitrogen Oxide (NOx)
Carbon Monoxide (CO)
750 & 1000
5.5% ± 0.2%
≤100 ppm
<150 ppm
1500
5.2% ± 0.2%
≤100 ppm
<150 ppm
2000
6.0% ± 0.2%
≤100 ppm
<150 ppm
2500
5.0% ± 0.2%
≤100 ppm
<150 ppm
3000
5.2% ± 0.2%
≤100 ppm
<150 ppm
4000
4.5% ± 0.2%
≤100 ppm
<150 ppm
5000N
4.5% ± 0.2%
≤100 ppm
<150 ppm
5000
5.5% ± 0.5%
≤100 ppm
<150 ppm
6000
5.0% ± 0.5%
≤100 ppm
<150 ppm
NOTE:
These instructions assume that the inlet air temperature is between 50°F and 100°F (10°C
37.8°C). If NOx readings exceed the target values in Table 4-4, above, or Table 4-6, below,
increase the O
2
level up to 1% higher than the Target value. You must then record the
increased O
2
value on the Combustion Calibration sheet.
22. On Benchmark 3000 - 6000 units only, record the manifold (downstream) gas pressure at
100%. This value will be used in Section 5.2.2: Low Pressure Gas Test, and Section 5.3.2:
High Pressure Gas Test.

Table of Contents

Other manuals for Watts AERCO Benchmark BMK5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Watts AERCO Benchmark BMK5000 and is the answer not in the manual?

Watts AERCO Benchmark BMK5000 Specifications

General IconGeneral
BrandWatts
ModelAERCO Benchmark BMK5000
CategoryBoiler
LanguageEnglish

Related product manuals