WM 68331 04/2021
LIFE-BASE EN 151
8 Hygienic reprocessing
6. If available: Remove the protective bags and accessories bags
from the portable unit (see "7 Disassembly", page 110).
7. Remove the devices from the portable unit (see "7.5 Removing
the device from the portable unit", page 123).
Result The parts have been prepared for hygienic reprocessing.
8.3 Cleaning parts manually
Requirement • The parts are visibly soiled.
• The hygienic reprocessing is prepared (see "8.2 Preparing
hygienic reprocessing", page 150).
1. Refer to the cleaning and disinfection plan for approved parts,
approved agents and dosages (see "8.7 Cleaning and
disinfection plan", page 155).
2. Prepare the cleaning solution in accordance with the
specifications of the cleaning agent manufacturer.
3. To remove all visible soiling: Thoroughly brush down parts
using a standard soft brush that is suitable for use on plastics
and that has been wetted with the cleaning agent.
When doing so, note: Uneven surfaces and grooves (e.g.,hook
and loop fasteners, zippers, belt buckles) must be particularly
well wetted with the disinfectant.
4. Wipe parts down with a clean, lint-free disposable cloth that
has been moistened with cleaning solution.
When doing so, note:
• A new disposable cloth must be used each time for
cleaning.
• All surfaces must be wiped down carefully.
• All surfaces must be wetted with cleaning solution.
• Ensure compliance with the cleaning cycles specified in
the cleaning and disinfection plan.
• Uneven surfaces and grooves must be wiped particularly
thoroughly.