8 Hygienic reprocessing
8.7 Cleaning and disinfection plan
Carry out hygienic reprocessing according to the following table
after every use:
Level of
disinfection
Description
High Level
Disinfection
A germ-killing process with a sterilizing agent
under non-sterile conditions. The process kills
all types of microbial life with the exception of
a large number of bacterial spores.
Intermediate Level
Disinfection
A germ-killing process using a disinfectant that
kills viruses, mycobacteria, fungi and vegetative
bacteria, but not bacterial spores.
Low Level
Disinfection
A germ-killing process using a disinfectant that
kills some fungi, vegetative types of bacteria
and lipid viruses.
Part
Level of
disinfec-
tion
Manual
cleaning
Wipe disin-
fection
Immersion
disinfec-
tion
Mechani-
cal repro-
cessing
Steriliza-
tion
Portable unit and wall mounting
Portable unit
LIFE-BASE
Intermediate
Level
Disinfection
Wipe down
with
neodisher
®
MediClean
forte
(Dr. Weigert).
Dosage:
10 ml/l.
Treat all
surfaces at
least twice
until they are
visibly clean
Wipe down
all surfaces
once with
Incidin™
OxyWipe S
(Ecolab).
Exposure
time: 5 min
Not permitted Not permitted Not permitted
Wall moun-
ting BASE-
STATION

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann LIFE-BASE Series and is the answer not in the manual?

Weinmann LIFE-BASE Series Specifications

General IconGeneral
BrandWeinmann
ModelLIFE-BASE Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals