Page #156 background image
156 EN LIFE-BASE
WM 68331 04/2021
8 Hygienic reprocessing
The applicable instructions are those in the instructions for use
from the manufacturers of the individual components or parts.
Observe these instructions for use.
Power supply
Charging
interface
Intermediate
Level
Disinfection
Wipe down
with
neodisher
®
MediClean
forte
(Dr. Weigert).
Dosage:
10 ml/l.
Treat all
surfaces at
least twice
until they are
visibly clean
Wipe down
all surfaces
once with
Incidin™
OxyWipe S
(Ecolab).
Exposure
time: 5 min
Not permitted Not permitted Not permitted
Bags and straps
Protective
bags
Intermediate
Level
Disinfection
Wipe down
smooth
surfaces with
neodisher
®
MediClean
forte
(Dr. Weigert).
Dosage:
10 ml/l.
Treat all
surfaces at
least twice
until they are
visibly clean
Not permitted Not permitted
Wash
without
spinning in
a washing
machine.
Pre-washing
5 min at 20°C
Main washing
30 min at
30°C
Cleaning
agent:
Turbo Usona
(Ecolab),
Disinfectant:
Turbo Oxysan
Not permitted
Accessories
bags
Shoulder
straps
Tension
belts
Devices and accessories
Follow the instructions for use for the devices and accessories.
Part
Level of
disinfec-
tion
Manual
cleaning
Wipe disin-
fection
Immersion
disinfec-
tion
Mechani-
cal repro-
cessing
Steriliza-
tion

Table of Contents