82 EN LIFE-BASE
WM 68331 04/2021
5 Function check
5.1 Intervals
5.2 Checking the portable unit visually
1. Check the portable unit for external damage.
2. If available: Check the charging interface and
connection line for:
Damage
Wear
Exposed wires
Bent connection lines
3. If available: Supply the portable unit with voltage via
the charging interface.
When doing so, note:
The LED for the external power supply must light up.
The battery status indicator must light up or flash.
4. If parts are damaged: Either have the portable unit repaired
by WEINMANN Emergency or replace it.
Result The portable unit has been checked visually.
Part Interval
Portable unit including all attached
devices
Before each use
After each assembly
After each hygienic reprocessing
At least every 6 months