134 EN LIFE-BASE
WM 68331 04/2021
7 Disassembly
Disassembling the monitor/defibrillator
Applies to • MEDUCORE Standard
2
• MEDUCORE Standard
Required auxiliary
equipment
Screwdriver with crosshead PH1
Requirement Disassembly is prepared (see "7.5.1 Preparing disassembly",
page 123).
1. With LIFE-BASE 1 NG XL:
• Undo the knurled screws.
• Pull out the hangers.
• Tighten the knurled screws.
2. Place the portable unit and device with the control panel
membrane facing downwards.