106 EN MEDUMAT Standard
2
WM 68071 05/2021
4 Preparation and operation
4.18 Using the simulation mode
The device features a simulation mode with which settings can
be simulated.
Requirement The device is switched off.
1. Switch on the device and during start-up press the right-hand
and left-hand navigation knobs at the same time.
The device switches to simulation mode.
The words Simulation Mode! flash in the display.
Risk of injury from confusing simulation mode with the
device's normal mode!
The only difference between the two modes are the words
Simulation Mode! on the display, so it is easy to confuse the
two if overlooked. This can put the patient at risk.
Do not use the simulation mode when the device is in use.
Always switch off and restart the device after using the
simulation mode.