WM 68071 05/2021
MEDUMAT Standard
2
EN 185
8 Hygienic reprocessing
• Wet any uneven surfaces and grooves (e.g., adjusting
knob, connection for ventilation tube) sufficiently with
the disinfectant.
• With a change of filter: Wipe disinfect the filter
compartment.
4. Allow the parts to dry fully.
5. Check the parts for residue and any remaining dirt.
6. If visible dirt remains: Repeat wipe disinfection.
Result The parts are wipe disinfected.
8.7 Immersion disinfecting parts
Requirement The parts intended for immersion disinfection have been manually
cleaned (see "8.11 Cleaning and disinfection plan", page 196).
1. The parts approved for immersion disinfection can be found in
the cleaning and disinfection plan (see "8.11 Cleaning and
disinfection plan", page 196).
2. The disinfectants, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page 196).
3. Prepare the immersion disinfection solution in accordance with
the specifications of the disinfectant agent manufacturer.
4. Immerse the parts in the immersion disinfection solution.
When doing so, note:
• All cavities must be filled.
• There must not be any air bubbles.
• All surfaces must be wetted.
• Swirl the parts in the immersion disinfection solution to
coat all surfaces and any cavities completely.
• Ensure compliance with the exposure time specified in
the cleaning and disinfection plan.