7 Description of the modes
7.4 Additional functions
7.4.1 CPR mode
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
2
monitoring are output in the additional functions CPR
and RSI. This can injure the patient.
Notify the user if the alarms in the additional functions CPR
and RSI have been deactivated.
Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
Description
Abbreviation CPR Manual CPR IPPV CPR CCSV
Long form Cardiopulmonary Resuscitation
Type Volume-controlled Pressure-controlled
Requirement None None
•Flow
measurement +
ASB option is
activated
CCSV option is
activated
Ventilation parameters
Left-hand navigation
knob
Vt Vt Trigger
Central navigation
knob
-Freq.
Right-hand navigation
knob
•CPR
mode
pMax
•CPR
mode
•Pause
•pMax
•PEEP
•CPR mode
•pInsp
•PEEP

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals