EasyManua.ls Logo

Weinmann MEDUMAT Standard2 - After Use

Weinmann MEDUMAT Standard2
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
94 EN MEDUMAT Standard
2
WM 68071 05/2021
4 Preparation and operation
Air Mix operation:
Result The operating time has been calculated.
4.12 After use
1. Detach the breathing circuit from the ventilation mask or
endotracheal tube.
2. If necessary: Dispose of the ventilation mask or endotracheal
tube.
3. If necessary: Disconnect the breathing circuit from the device.
4. If necessary: Dispose of the disposable breathing circuit.
5. If necessary: Take a new disposable breathing circuit.
6. If necessary: Change hygiene filter.
7. Hygienically reprocess the device, components and accessories
(see "8 Hygienic reprocessing", page 171).
8. If necessary: Take a new ventilation mask or new endotracheal
tube.
9. If necessary: Stow the components and accessories away on
the portable unit.
10. If necessary: Store the device, components and accessories
(see "12 Storage", page 227).
Example
Oxygen supply 2000 l
Vt 500 ml
f 12 min
-1
Time
634 min = 10 h 34 min
or
63 min per liter of cylinder volume
Time (min)=
Oxygen supply (I) x 2
Vt (l) x f (min
-1
) + 0.3 l/min

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Related product manuals