WM 68071 05/2021
MEDUMAT Standard
2
EN 201
9 Function check
5. Check the system for leaks (see "9.6 Checking the system for
leaks", page 210).
If necessary: Rectify any leaks in the system (see "9.7 Rectifying
leaks in the system", page 211).
Result The function check is ready.
9.3 Performing a function check
You can perform the function check with the following test lungs:
• Testing bag WM 1454
• EasyLung for WEINMANN Emergency WM 28625
Requirement • The device is disconnected from the patient.
• A fully charged battery is inserted in the device.
• The hygiene filter is inserted in the device.
• The protective cap has been removed from the end of
the breathing circuit.
Risk of injury from incorrect test lungs!
Test lungs other than those named here may not reliably
detect errors and thus distort the result of a function check.
This can injure the patient.
Only use the test lungs named here.
Risk of injury due to a connection between the device and
the patient during the function check!
A connection between the device and the patient during the
function check can result in barotrauma and injury to the patient.
Always disconnect any connection between the device and
the patient for the function check.
Material damage from touching the contacts in the battery
compartment!
Touching the contacts and the ingress of liquids can damage
the device.
Only operate the device with the battery inserted.
Do not touch the contacts in the battery compartment.
Clean the battery compartment carefully so that no liquids
enter the device.