11 Maintenance
11.5 Replacing the hygiene filter
Requirement The device is switched off.
1. Wipe-disinfect the outside of the hygiene filter and the device.
2. Pull the hygiene filter out of the filter compartment of the
device.
3. Dispose of the hygiene filter along with the filter cassette
(see "13.4 Hygiene filter", page 230).
4. With the filter side facing forwards, slide the hygiene filter into
the device's filter compartment until the hygiene filter is flush
with the device.
5. Perform a function check (see "9.3 Performing a function
check", page 201).
Result The hygiene filter has been replaced.