WM 68071 05/2021
MEDUMAT Standard
2
EN 51
4 Preparation and operation
4.4 Connecting components
4.4.1 Inserting the hygiene filter
1. Check the hygiene filter for external damage.
If necessary: Replace the hygiene filter.
2. With the filter side facing forwards, slide the hygiene filter into
the device's filter compartment until the hygiene filter is flush
with the device.
3. Perform a function check (see "9.3 Performing a function
check", page 201).
Result The hygiene filter has been inserted.
4.4.2 Connecting the breathing circuit
Risk of injury posed by ventilation with inhalation mask,
tube or nasal cannula!
Ventilation with an inhalation mask, tube or nasal cannula
connected may cause an injury to the patient.
Do not use an inhalation mask, tube or nasal cannula
for ventilation.
Recommendation: Always use a breathing system filter
for ventilation.