200 EN MEDUMAT Standard
2
WM 68071 05/2021
9 Function check
9 Function check
9.1 Intervals
Carry out a function check at regular intervals:
9.2 Preparing for the function check
1. Check battery status: The battery must be fully charged.
If necessary: Charge or replace the battery.
2. Check the following parts for external damage:
Device
Plug and cable
Breathing circuit
Accessories
If necessary: Replace parts.
3. Check the patient valve of the breathing circuit
(see "9.5 Testing the reusable breathing circuit", page 210).
If necessary: Replace the breathing circuit.
4. Check the oxygen level in the oxygen cylinder.
If necessary: Change the oxygen cylinder.
Part concerned Interval
Device
Before each use
After each hygienic reprocessing
After each repair
Breathing circuit (reusable
breathing circuit)
Before each use
After each hygienic reprocessing
After each disassembly
At least every 6 months