WM 68071 05/2021
MEDUMAT Standard
2
EN 25
3 Description
3.2 Control panel
3-2 Controls
1
2
4
3
5
6
7
89
No. Designation Description
1 Line power indicator
Steady green light: Indicates that the device is
connected to line power.
Not illuminated: The device is operating on
battery power and not on line power.
or
The device is in NVG mode.
2 Battery status indicator
Steady green light: The battery is full or is not
being charged because it is outside the charging
temperature range.
Flashing green light: The battery is being charged.
Steady red light: The battery is defective or not in
the device.
Not illuminated: The device is operating on
battery power and not on line power.
or
The device is in NVG mode.