176 EN MEDUMAT Standard
2
WM 68071 05/2021
8 Hygienic reprocessing
8.4 Disassembly of the reusable
breathing circuit
The images in this subchapter show all the possible components of
the reusable breathing circuit. Depending on the system type, your
reusable breathing circuit may not include certain components
(see "3.5.2 Reusable breathing circuit and disposable breathing
circuit", page 35).
Requirement The device is disconnected from the breathing circuit.
The patient is disconnected from the breathing circuit.
1. Open the protective sleeve.
2. Open the hook and loop fasteners in the protective sleeve.
3. If available: Detach the water filter from the measuring circuit
connector.
4. If available: Detach the water filter from the CO
2
measuring
tube.
Material damage from pulling out the hoses/tubes incorrectly!
The hoses/tubes may be damaged if pulled out incorrectly.
Hold the end of the hose/tube when pulling out!