12 Storage
12.4 Storing the battery
Requirement The device and the battery have been cleaned and disinfected
(see "8.3 Preparing hygienic reprocessing", page 175).
The battery is fully charged.
If available: The replacement battery is fully charged.
1. Insert the battery in the battery compartment and store the
device in a dry place.
or
Store the battery in a dry place outside of the device.
2. Charge battery at regular intervals:
Result The battery is stored in a dry place and is ready for use.
Material damage due to prolonged storage of the battery
without recharging!
Storing the battery for a prolonged period of time without
recharging can result in the rapid shutdown of and irreparable
damage to the battery.
When the battery is stored in the device without a power
connection: Charge battery every 3 months.
If the battery is not stored in the device: Charge battery with
SN < 20000 every 5 months and battery with SN 20000
every 9 months.
Type of storage Charging interval
In device without power connection Every 3 months
Not in device
Battery with SN < 20000
every 5 months
Battery with SN 20000
every 9 months

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals