9 Function check
18. Pull the test lung from the breathing circuit.
Result The function check is complete.
9.4 Failed function check
Requirement The function check ended with Device is not ready for use.
1. Check the components named in the instructions on
the display and replace if necessary.
2. Repeat the function check.
3. If the function check ends with Device is not ready for use
again: Contact your authorized dealer or
WEINMANN Emergency.
Risk of injury from improperly removed testing bag!
If the testing bag is removed improperly, the connector of the
testing bag may remain on the breathing circuit. The resulting
increase in inspiratory airway resistance can injure the patient.
When disassembling always pull the testing bag off at the
connector.
Risk of injury due to inoperational device!
Operation of the device after a failed function check may result
in injury to the patient.
Only operate the device after it passes the function check.
Precise information on the individual tests in the function check
can be found in the file fcheck (see "14.3.1 Recorded function
checks", page 259).