9 Function check
14. Proceed with the device according to the following table:
15. Press the navigation knob finish.
16. Switch off the device.
17. Close the oxygen cylinder.
Display Meaning Action
Device is ready for use Function check passed. Use device without restriction.
• Device is ready for use
• Service due in %1 days
• Function check passed
• ≤ 60 days until expiry of
the maintenance interval
• Device can be used
without restrictions
until expiry of the
maintenance interval.
• To continue using the
device without
restrictions: Contact
WEINMANN Emergency
or a technician
authorized by
WEINMANN Emergency
in good time for
maintenance.
• Device is ready for use
• Service due
• Maintenance symbol is
flashing in the display
(only in the start menu)
• Function check passed
• Maintenance interval
expired
To continue using the device
without restrictions: Contact
WEINMANN Emergency or
a technician authorized by
WEINMANN Emergency for
maintenance.
• Device is ready for use
• Check or replace the
FlowCheck sensor
• Maintenance symbol is
flashing in the display
(only in the start menu)
• Function check passed
• Total useful life of the
FlowCheck sensor has
been exceeded
To continue using the device
without restrictions: Check or
replace the FlowCheck sensor.
Device is not ready for use Function check failed.
Take action (see "9.4 Failed
function check", page 209).
The message Check or replace the FlowCheck sensor can also
appear with a notice that maintenance is required.