196 EN MEDUMAT Standard
2
WM 68071 05/2021
8 Hygienic reprocessing
8.11 Cleaning and disinfection plan
Carry out hygienic reprocessing according to the following table
after every use:
8.11.1 Device and accessories
Part Manual cleaning
Wipe
disinfection
Immersion
disinfection
Mechanical
reprocessing
(disinfection)
Device Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dosage: 10 ml/l,
Duration: treat all
surfaces at least
twice until visibly
clean
Wipe down with
Incidin™ OxyWipe
S (Ecolab).
Exposure time:
5 min
Not permitted Not permitted
12 V cable
Charging adapter
Battery
Power supply unit
Testing bag
Hygiene filter
Disposable, do not reuse but dispose of properly after transportation of an infected
patient or upon expiry of the filter's service life (see 11.2, p. 223) (see 13, p. 230)
Inhalation adapter Disposable, do not reuse. Dispose of correctly (see 13, p. 230).
Ventilation mask
Observe the instructions for use from the manufacturer.
Endotracheal tube
Breathing system
filter
Portable unit and
protective bags
See instructions for use for LIFE-BASE portable unit
OXYWAY Pressure
reducer
See instructions for use for OXYWAY pressure reducer
Low-pressure hoses See instructions for use for low-pressure hose LPH

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals