WM 68071 05/2021
MEDUMAT Standard
2
EN 197
8 Hygienic reprocessing
8.11.2 Breathing circuits
Part Manual cleaning
Wipe
disinfection
Immersion
disinfection
Mechanical
reprocessing
(Disinfection)
Reusable breathing circuit
Ventilation hose
Immerse in and
clean with
neodisher
®
MediClean forte
(Dr. Weigert).
Dosage:
10 ml/l,
Exposure time:
10 min
Not permitted
Immerse in
gigasept
®
FF neu
(Schülke)
Dosage:
50 ml/l
Exposure time:
15 min
Cleaning:
neodisher
®
MediClean forte
(Dr. Weigert):
0.5 %, 55 °C,
10 min
Thermal
disinfection:
90 °C, 5 min
(corresponding to
A0 value 3000)
Steam sterilization:
5 min, 134 °C
Patient valve
Elbow
Protective cap
FlowCheck sensor
(reusable)
Connector with
CO
2
connection
Reusable
measuring
circuit
Immerse in and
clean with
neodisher
®
MediClean forte
(Dr. Weigert).
Dosage:
10 ml/l,
Exposure time:
10 min
Not permitted
Immerse in
gigasept
®
FF neu
(Schülke)
Dosage:
50 ml/l
Exposure time:
15 min
Cleaning:
neodisher
®
MediClean forte
(Dr. Weigert):
0.5 %, 55 °C,
10 min
Thermal
disinfection:
90 °C, 5 min
(corresponding to
A0 value 3000)
CO
2
measuring
tube (only with
Capnography
option)
Disposable, do not reuse. Dispose of correctly (see 13, p. 230).
Water filter
(only with
Capnography
option)
Disposable, do not reuse. Dispose of correctly (see 13, p. 230).