EasyManuals Logo

Weinmann MEDUMAT Standard2 User Manual

Weinmann MEDUMAT Standard2
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
WM 68071 05/2021
MEDUMAT Standard
2
EN 195
8 Hygienic reprocessing
11. If available: Connect the CO
2
measuring tube with water filter
to the measuring circuit connector.
12. If available: Seal the end of the reusable breathing circuit
closest to the patient with a protective cap.
13. Place all the hoses/tubes and one of the connection lines in
the protective sleeve.
14. Close the hook and loop fasteners in the protective sleeve
around the hoses/tubes and connection line.
15. Close the protective sleeve zipper.
Result The reusable breathing circuit is assembled.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals