8 Hygienic reprocessing
11. If available: Connect the CO
2
measuring tube with water filter
to the measuring circuit connector.
12. If available: Seal the end of the reusable breathing circuit
closest to the patient with a protective cap.
13. Place all the hoses/tubes and one of the connection lines in
the protective sleeve.
14. Close the hook and loop fasteners in the protective sleeve
around the hoses/tubes and connection line.
15. Close the protective sleeve zipper.
Result The reusable breathing circuit is assembled.