WM 68071 05/2021
MEDUMAT Standard
2
EN 199
8 Hygienic reprocessing
Disposable breathing circuit
Disposable
breathing circuit
Disposable, do not reuse. Dispose of correctly (see 13, p. 230).
Disposable
measuring circuit
Hose clips
FlowCheck sensor
(disposable)
Water filter
CO
2
measuring
tube (only with
Capnography
option)
FlowCheck sensor
connection line
Wipe down with
neodisher
®
MediClean forte
(Dr. Weigert).
Dosage: 10 ml/l,
Duration: treat all
surfaces at least
twice until visibly
clean
Wipe down
with Incidin™
OxyWipe S (Ecolab).
Exposure time:
5 min
Not permitted Not permitted
FlowCheck sensor
connection line
with MEDUtrigger
MEDUtrigger with
connection line
Part Manual cleaning
Wipe
disinfection
Immersion
disinfection
Mechanical
reprocessing
(Disinfection)
The applicable instructions are those in the instructions for use
from the manufacturers of the individual components or parts.
Observe these instructions for use.