186 EN MEDUMAT Standard
2
WM 68071 05/2021
8 Hygienic reprocessing
5. After the specified exposure time, rinse the parts for at least
10 s and textiles for at least 2 min with drinking-water quality
in order to remove all of the disinfectant residue.
6. Allow the parts to dry fully.
7. Check the parts for residue and any remaining dirt.
8. If dirt is visible: Repeat the cleaning and disinfection.
Result The parts are immersion disinfected.
8.7.1 Immerse disinfecting the reusable measuring
circuit
The principle described only applies to the following parts of the
reusable measuring circuit:
Pressure measuring tube
PEEP control tube
Measuring circuit connector
Requirement The reusable measuring circuit is disconnected from the
reusable breathing circuit (see "8.4 Disassembly of the
reusable breathing circuit", page 176).
The reusable measuring circuit has been cleaned manually.
1. The disinfectants, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page 196).
2. Prepare the immersion disinfection solution in accordance with
the specifications of the disinfectant agent manufacturer.
3. Connect a sterile disposable syringe (20 ml) to one end of a
tube.
Risk of infection and contamination due to insufficient
hygienic reprocessing of the measuring circuit!
Rinsing the measuring circuit in alternating directions does not
guarantee any bacteria reduction and may injure the patient.
Only rinse measuring tubes in one direction.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals