112 EN MEDUMAT Standard
2
WM 68071 05/2021
5 User menu
5.3.2 Alarm limits
5-3 Submenu Alarm limits
Here you can set the alarm limits.
You can also open the alarm limit menu by keeping the alarm mute
button depressed.
Risk of injury due to alarm limits which are too high or too
low!
Alarm limits which are either too high or too low can prevent the
device from emitting an alarm, thereby putting the patient at risk.
Always set alarm limits which have been adapted to the
patient.
Alarm Setting range
Activating automatic limits The device sets the alarm limits for
the alarms relating to respiratory
physiology automatically.
The deviation is 10 %, 20 % or 30 %
from the ventilation values at the time
of activation. The automatic alarm
limits are set to ± 30 % on delivery.
Autom. limits
MVe (only with Flow
measurement + ASB option)
1 l to 160 l