14 EN MEDUMAT Standard
2
WM 68071 05/2021
2 Safety
Risk of injury as a result of pressure variations during use in
combination with devices from the WEINMANN Emergency
MODUL range!
If the device is used together with devices from the
WEINMANN Emergency MODUL range, the flow used by devices
from the WEINMANN Emergency MODUL series may cause
pressure variations in the device. This can injure the patient.
Only use the device and devices from the
WEINMANN Emergency MODUL range in combinations
approved by WEINMANN Emergency.
Prevent interference between the devices!
Electrical devices which are operated directly next to or on top of
each other can cause mutual interference to functionality.
Portable high-frequency communication devices in the direct
vicinity of the device can also influence the functioning of the
device.
Do not stack the device with other electrical devices.
Do not operate the device directly next to other electrical
devices. Exception: Other WEINMANN Emergency devices
which have been tested and shown to guarantee interference-
free operation with the adjacent device. A list of other devices
is available on request.
If stacking or operation in the immediate vicinity cannot be
avoided: Closely monitor the functioning of all affected
medical electrical devices and do not use if functions are
disrupted.
With portable RF communication devices, maintain a minimum
distance of 30 cm to the device and accessories. Examples:
Wireless device, mobile telephone.
Notice Damage to the device caused by ingress of liquids!
The device is rated IP54 (splash-proof). This only applies when the
battery is located in the battery compartment, the SD card
compartment is closed, there is a filter in the filter compartment
and there is a breathing circuit connected. Ingress of liquids may
damage the device, components and accessories.
Do not immerse the device, components or accessories in
liquids.
Clean the battery compartment carefully so that no liquids
enter the device.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals