142 EN MEDUMAT Standard
2
WM 68071 05/2021
7 Description of the modes
• If the Flow measurement + ASB option is not activated:
With a set PEEP value > 0 mbar, the patient must create an
underpressure of at least -1.2 mbar below the set PEEP value
through his/her spontaneous respiratory effort in order to
initiate an inspiratory trigger in the device.
InTr level (only with
activated Flow
measurement + ASB
option)
- Inspiratory trigger (three-level setting)
ExTr level (only with
activated Flow
measurement + ASB
option)
- Expiratory trigger (three-level setting)
InTr (only with
activated Flow
measurement + ASB
option)
l/min Inspiratory trigger (setting with units)
ExTr (only with
activated Flow
measurement + ASB
option)
% Flow
max
Expiratory trigger (setting with units)
I:E - Inspiration to expiration ratio
(only with activated
Flow
measurement + ASB
option)
- Pressure increase time
Trigger window (only
in aPCV mode with
activated Pressure
controlled ventilation
modes option)
% Te Trigger window
Ventilation
parameters
Unit Description