7 Description of the modes
7.3.3 CPAP mode
The set value CPAP/PEEP is used to increase the pressure level of
respiration in order to raise the functional residual capacity (FRC)
of a spontaneously breathing patient. The patient is able to
breathe spontaneously without any restriction at the set pressure
level. The CPAP mode is used exclusively on patients with adequate
spontaneous breathing.
In principle, the pressure is set at the end of expiration (PEEP).
The set maximum ventilation pressure (pMax) ensures the safety of
the patient.
Description
Abbreviation CPAP
Long form Continuous Positive Airway Pressure
Type Pressure-controlled
Requirement None
Ventilation parameters
Left-hand navigation knob -
Central navigation knob -
Right-hand navigation knob
•PEEP
pMax
Emergency mode
CPAP
Pressure
Time
1/Freq. (spontaneous) 1/Freq. (spontaneous)
pMax
CPAP / PEEP
CPAP

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals