7 Description of the modes
7.3.6 PCV mode
Description
Abbreviation PCV
Long form Pressure Controlled Ventilation
Type Pressure-controlled
Requirement
Flow measurement + ASB option
is activated
Pressure controlled ventilation
modes option is activated
Curve display option is activated
Ventilation parameters
Left-hand navigation knob pInsp
Central navigation knob Freq.
Right-hand navigation knob
PEEP
pMax
I:E
Emergency mode
Pressure
Time
PCV
pMax
1/Freq. (set)
PEEP
1/Freq. (set)
dP
dT
Pressure
ramp
pInsp