7 Description of the modes
If this inspiratory pressure is achieved (pInsp = pMax - 5 mbar),
the device administers as much volume as possible. If this
volume deviates from the set tidal volume, the device triggers
the medium-priority alarm Vt not achievable.
7.3.8 SIMV mode
Description
Abbreviation SIMV
Long form
Synchronized Intermittent Mandatory
Ventilation
Type Volume-controlled
Requirement SIMV option is activated
Ventilation parameters
Left-hand navigation knob Vt
Central navigation knob Freq.
Right-hand navigation knob
•PEEP
• pMax
•I:E
• Emergency mode
SIMV
1/Freq. (set)
Mechanical
ventilation Synchronized
mechanical
ventilation
1/Freq. (current)
Trigger time slot
'T
Spontaneous respiration
Patient
trigger
PEEP
p Max