158 EN MEDUMAT Standard
2
WM 68071 05/2021
7 Description of the modes
7.3.10 S-IPPV mode
Risk of hyperventilation!
When using the S-IPPV mode, the CO
2
concentration in the
patient's blood can drop and injure the patient.
Monitor the patient continuously.
Risk of air trapping!
When using the S-IPPV mode, air can become trapped in the
patient's lung. This results in a reduced gas exchange and can
injure the patient.
Monitor the airway pressure continuously.
Risk of intrinsic PEEP!
An expiration that is too short can cause the pressure to increase
slowly at the end of the expiration and injure the patient.
Set the pressure limitation correctly.
Monitor the patient continuously.
Description
Abbreviation S-IPPV
Long form
Synchronized Intermittent Positive
Pressure Ventilation
Type Volume-controlled
Requirement S-IPPV option is activated
Ventilation parameters
Left-hand navigation knob Vt
Central navigation knob Freq.
Right-hand navigation knob
•PEEP
pMax
InTr (only with activated Flow
measurement + ASB option)
•I:E
Emergency mode

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals