8 Hygienic reprocessing
Disruption or failure of therapy due to unsuitable cleaning
agents and disinfectants!
The use of incorrect cleaning agents and disinfectants can
cause the device to malfunction. This can result in serious or
life-threatening injury to the patient.
Never clean the device and accessories with bleach, bleach
solutions or phenolic compounds.
Only use the cleaning agents and disinfectants recommended
in these instructions for use (see "8.11 Cleaning and
disinfection plan", page 196).
Loss of mechanical or electrical safety resulting from
reprocessing of the device and accessories with unsuitable
cleaning agents and disinfectants!
Using incorrect cleaning agents and disinfectants might cause
damage to the surface of the device and/or accessories, as well as
impairing electrical and insulating properties. This can result in
severe or life-threatening injury to the user and patient.
Only use the cleaning agents and disinfectants recommended
in these instructions for use (see "8.11 Cleaning and
disinfection plan", page 196).
Disruption or failure of therapy due to liquid in the breathing
circuit after hygienic reprocessing!
Drips in the reusable measuring circuit or FlowCheck sensor can
distort the measurement results. This can result in serious or
life-threatening injury to the patient.
Following hygienic reprocessing of the breathing circuit,
allow all breathing circuit components to dry thoroughly.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals