182 EN MEDUMAT Standard
2
WM 68071 05/2021
8 Hygienic reprocessing
5. If the parts need to be wiped down acc. to the cleaning and
disinfection plan: Wipe parts down with a lint-free clean cloth
that has been moistened with cleaning solution at least twice.
When doing so, note:
Use a new cloth for each cleaning procedure.
Carefully wipe down all the surfaces.
All surfaces must be wetted with cleaning solution.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
Rewipe uneven surfaces and grooves in particular.
6. If the parts need to be immersed acc. to the cleaning and
disinfection plan: Immerse parts in the cleaning solution.
When doing so, note:
Swirl the parts in the cleaning solution to coat all surfaces
and any cavities completely.
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
7. If visible dirt remains: Repeat the manual cleaning.
8. Rinse the parts immersed in the cleaning solution for at least
10 s and textiles for at least 2 min thoroughly with drinking-
water quality.
9. Wipe down the remaining parts with a damp cloth twice in
order to remove cleaning agent residue.

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals