188 EN MEDUMAT Standard
2
WM 68071 05/2021
8 Hygienic reprocessing
2. Place the parts in a washer and disinfector.
When doing so, note:
Ensure compliance with the exposure time specified in the
cleaning and disinfection plan.
Connect the hoses/tubes with the washer and disinfector.
All parts and lumens must be completely flushed.
The water must be able to run away.
3. Add cleaning agents as per the instructions for use for the
washer and disinfector.
4. If necessary: Add neutralizing agent as per the instructions for
use for the washer and disinfector.
5. Start machine reprocessing cycle according to the following
process, e.g., cycle 12 of the washer and disinfector
G 7836 CD from Miele.
6. Check the parts for residue and any remaining dirt.
7. If visible dirt remains: Repeat the machine cleaning and
disinfection.
Result The parts are machine cleaned and disinfected.
Step Process Temperature
Exposure
time
Cleaning
agent/
disinfectant
1 Initial cleaning Cold water 1 min
2 Main cleaning 55 °C 10 min
neodisher
®
MediClean forte
(Dr. Weigert):
0.5 %
3 Neutralization Cold water 2 min
4 Rinsing Cold water 3 min
5
Thermal
disinfection
90 °C 5 min
Demineralized
water
6 Drying 100 °C 25 min

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals