204 EN MEDUMAT Standard
2
WM 68071 05/2021
9 Function check
6. Check the alarm system:
The alarm light must flash red.
The device must emit at least one audible alarm in alarm
system test 1 and alarm system test 2.
7. If the alarm system is functioning: Press the navigation knob
yes each time.
8. If the alarm system is not functioning: Press the navigation
knob no.
If the software of the FlowCheck sensor connection line/
FlowCheck sensor connection line with MEDUtrigger is not
up-to-date, the device updates the software before the key
function check starts.