2 Safety
• Repairs, servicing and maintenance should only be carried
out by the manufacturer, WEINMANN Emergency, or by a
technician expressly authorized by WEINMANN Emergency.
The device's service and repair instructions are available to
the latter; these contain all the information required.
• Only have modifications to the device carried out by the
manufacturer, WEINMANN Emergency, or by a technician
expressly authorized by WEINMANN Emergency.
• Any constructive changes made to the device may put the
patient and the user at risk and are not permitted.
• The device is protected against unauthorized access by means
of a colored security seal on the rear of the housing. Please
note that any damage to the security seal voids any warranty
claims.
• Please observe the section on hygienic reprocessing in order to
avoid infection or bacterial contamination (see "8 Hygienic
reprocessing", page 171).
• Also observe the respective instructions for use for the device,
the components and the accessories.
• Always carry out a function check before using the device
(see "9 Function check", page 200).
• As the user, always remain in the direct vicinity of the device
and patient.
• Always have an alternative respiration option in addition to
the ventilator on hand. An alternative respiration option is,
for example, a resuscitator for manual ventilation.
• The power supply unit/charger is not intended for use in
vehicles or outdoors. Only use the power supply unit/charger
in closed rooms and observe the technical data
(see "14.1 Technical data", page 231).
• Risks due to software errors have been minimized by means
of extensive qualification measures.