EasyManua.ls Logo

Weinmann MEDUMAT Standard2 - Page 236

Weinmann MEDUMAT Standard2
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
236 EN MEDUMAT Standard
2
WM 68071 05/2021
14 Appendix
(1)
BTPS (Body Temperature and Pressure, saturated): Volume at current
ambient pressure and 37 °C, with 100 % saturated gas
Trigger window for mandatory
mechanical breaths
20 % of Te (SIMV, SIMV + ASB, BiLevel + ASB and PRVC + ASB)
0 % to 100 % of Te (S-IPPV)
Trigger window for ASB mechanical
breaths
0 % to 100 % Te
Trigger window aPCV
(only with Pressure controlled
ventilation modes option)
0 % to 100 % Te (adjustable)
Volume monitoring
(1)
Measurement range: 40 ml to 8000 ml
Tolerance: ± 15 %
Airway pressure monitoring
Measurement range: -5 mbar to +80 mbar
Tolerance: ± 3 mbar
Oxygen concentration:
Air Mix operation
No Air Mix operation
See "14.1.8 Oxygen concentration with Air Mix operation",
page 251.
100 % oxygen
Concentrator oxygen (90 % to 96 % oxygen)
Pressurized gas thread External thread G 3/8
Connection for ventilation hose WEINMANN Emergency-specific
Patient valve connections WEINMANN Emergency-specific
Service life of hygiene filter 24 h in Air Mix operation or 6 months
Efficiency of hygiene filter > 99 % for viruses and bacteria (particle size > 27 nm)
Specification Device
Subject to alterations in design.