14 Appendix
Calibration of the CO
2
module
The CO
2
module is only calibrated within the scope of the 2-year
maintenance intervals and does not have to be calibrated by the
user. Neither calibration nor zero-point adjustment are performed
during operation. The CO
2
module features automatic adjustment
for barometric pressure. When used as intended, no specific
unwanted gases are emitted at the sampling point which would
have an influence on the gas measurement. The gas sample is
filtered and discharged into the environment via a pressure
equalization opening in the device housing. No measures are
required for the proper disposal of the gas sample.