14 Appendix
Guidelines and manufacturer's declaration – electromagnetic immunity
MEDUMAT Standard
2
is intended for operation in the electromagnetic environment specified below.
The customer or the user of the MEDUMAT Standard
2
device must ensure that it is also used in such
an environment.
Immunity tests
IEC 60601 test
level
Compliance
level
Electromagnetic environment –
guidelines
Conducted RF
bursts acc. to
IEC 61000-4-6
Emitted RF
bursts acc. to
IEC 61000-4-3
3 V
effective value
150 kHz to 80 MHz
Outside of the ISM
bands
a
6 V
effective value
150 kHz to 80 MHz
Within the ISM
bands
a
10 V/m
80 MHz to 2.7 GHz
3 V
6 V
30 V/m
Portable and mobile wireless devices should
not be used at closer distances to the
MEDUMAT Standard
2
device, including its
cables, than the recommended separation
distance calculated in accordance with the
corresponding equation for the frequency of
the transmitter.
Recommended separation distance:
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
where P is the maximum rated output power
of the transmitter in watts (W) as per
the manufacturer of the transmitter's
specifications and d is the recommended
separation distance in meters (m).
b
The field strength of fixed RF transmitters,
as determined by an electromagnetic site
survey
c
, should be lower than the compliance
level in each frequency range.
d
Interference is
possible in the vicinity of devices furnished
with the following pictogram.
d12P,=
d12P,=
d04P,=
d08P,=

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals