EasyManua.ls Logo

Weinmann MEDUMAT Standard2 - Page 254

Weinmann MEDUMAT Standard2
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
254 EN MEDUMAT Standard
2
WM 68071 05/2021
14 Appendix
Guidelines and manufacturer's declaration – electromagnetic immunity
MEDUMAT Standard
2
is intended for operation in the electromagnetic environment specified below.
The customer or the user of the MEDUMAT Standard
2
device must ensure that it is also used in such
an environment.
Immunity tests
IEC 60601 test
level
Compliance
level
Electromagnetic environment –
guidelines
Conducted RF
bursts acc. to
IEC 61000-4-6
Emitted RF
bursts acc. to
IEC 61000-4-3
3 V
effective value
150 kHz to 80 MHz
Outside of the ISM
bands
a
6 V
effective value
150 kHz to 80 MHz
Within the ISM
bands
a
10 V/m
80 MHz to 2.7 GHz
3 V
6 V
30 V/m
Portable and mobile wireless devices should
not be used at closer distances to the
MEDUMAT Standard
2
device, including its
cables, than the recommended separation
distance calculated in accordance with the
corresponding equation for the frequency of
the transmitter.
Recommended separation distance:
for 80 MHz to 800 MHz
for 800 MHz to 2.5 GHz
where P is the maximum rated output power
of the transmitter in watts (W) as per
the manufacturer of the transmitter's
specifications and d is the recommended
separation distance in meters (m).
b
The field strength of fixed RF transmitters,
as determined by an electromagnetic site
survey
c
, should be lower than the compliance
level in each frequency range.
d
Interference is
possible in the vicinity of devices furnished
with the following pictogram.
d12P,=
d12P,=
d04P,=
d08P,=

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Related product manuals