WM 68071 05/2021
MEDUMAT Standard
2
EN 257
14 Appendix
14.1.10 Factory settings for emergency modes and
ventilation modes
Ventilation
parameters
Adult Child Infant
Pressure ramp Averaging Averaging Averaging
ExTr
35 % Flow
Level 2
35 % Flow
Level 2
35 % Flow
Level 2
Freq. 10/min 20/min 30/min
I:E 1:1.7 1:1.7 1:2.8
InTr
5 l/min
Level 2
5 l/min
Level 2
5 l/min
Level 2
PEEP 0 mbar 0 mbar 0 mbar
PEEP CCSV 3 mbar 3 mbar –
pInsp 20 mbar 15 mbar 15 mbar
pInsp CCSV 60 mbar 40 mbar –
pMax 30 mbar 25 mbar 20 mbar
pMax CPR Contin. 30 mbar 25 mbar 20 mbar
pMax CPR Man. 20 mbar 20 mbar 20 mbar
pMax RSI 30 mbar 25 mbar 20 mbar
Trigger window 20 % Te 20 % Te 20 % Te
Vt 500 ml 200 ml 60 ml
ΔpASB 0 mbar 0 mbar 0 mbar
Alarm limit Adult Child Infant
MVe 6.5 l/min 5.2 l/min 2.3 l/min
MVe 3.5 l/min 2.8 l/min 1.3 l/min
f 13/min 26/min 39/min
Apnea 30 s 30 s 30 s
etCO
2
49 mmHg/6 vol% 49 mmHg/6 vol% 49 mmHg/6 vol%
etCO
2
27 mmHg/4 vol% 27 mmHg/4 vol% 27 mmHg/4 vol%