14.1.10 Factory settings for emergency modes and
ventilation modes
Ventilation
parameters
Adult Child Infant
Pressure ramp Averaging Averaging Averaging
ExTr
35 % Flow
Level 2
35 % Flow
Level 2
35 % Flow
Level 2
Freq. 10/min 20/min 30/min
I:E 1:1.7 1:1.7 1:2.8
InTr
5 l/min
Level 2
5 l/min
Level 2
5 l/min
Level 2
PEEP 0 mbar 0 mbar 0 mbar
PEEP CCSV 3 mbar 3 mbar
pInsp 20 mbar 15 mbar 15 mbar
pInsp CCSV 60 mbar 40 mbar
pMax 30 mbar 25 mbar 20 mbar
pMax CPR Contin. 30 mbar 25 mbar 20 mbar
pMax CPR Man. 20 mbar 20 mbar 20 mbar
pMax RSI 30 mbar 25 mbar 20 mbar
Trigger window 20 % Te 20 % Te 20 % Te
Vt 500 ml 200 ml 60 ml
ΔpASB 0 mbar 0 mbar 0 mbar
Alarm limit Adult Child Infant
MVe 6.5 l/min 5.2 l/min 2.3 l/min
MVe 3.5 l/min 2.8 l/min 1.3 l/min
f 13/min 26/min 39/min
Apnea 30 s 30 s 30 s
etCO
2
49 mmHg/6 vol% 49 mmHg/6 vol% 49 mmHg/6 vol%
etCO
2
27 mmHg/4 vol% 27 mmHg/4 vol% 27 mmHg/4 vol%

Table of Contents

Other manuals for Weinmann MEDUMAT Standard2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUMAT Standard2 and is the answer not in the manual?

Weinmann MEDUMAT Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUMAT Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals