WM 68071 05/2021
MEDUMAT Standard
2
EN 49
4 Preparation and operation
Device
If the device is switched on, you can see the battery status on
the display:
1 LED is flashing Battery status < 10 %
No LEDs are lit
Battery is deeply discharged.
Charge battery in the device
for 24 hours. After 24 hours:
• Green LED is lit: Battery
fully charged and ready
for use.
• Red LED or no LED is lit:
Battery defective.
Replace battery.
Red fault indicator
is lit
Battery defective.
Replace battery.
Status indicator Meaning
Battery status > 90 %
Battery status approx. 60 % - 90 %
Battery status approx. 40 % - 60 %
Battery status approx. 10 % - 40 %
Battery status < 10 %
• The last remaining segment in
the battery status symbol is red.
• The message Battery weak
appears in the display.
Status indicator Explanation Meaning