WM 68071 05/2021
MEDUMAT Standard
2
EN 69
4 Preparation and operation
4.7.3 Selecting a ventilation mode for a new patient
Requirement The device is switched off.
1. Switch on the device.
After the self-test, the device displays the start menu:
If the device was switched off for < 30 seconds (e.g., for a
battery replacement), a countdown in the Previous patient
field counts down 20 seconds.
If the device was switched off for > 30 seconds, the
countdown for the last patient is deactivated.
2. Select the New patient field.