WM 68071 05/2021
MEDUMAT Standard
2
EN 83
4 Preparation and operation
3. If you want chest compressions to be carried out automatically:
Switch to using the middle navigation knob.
or
If you want chest compressions to be carried out manually:
Switch to using the middle navigation knob.
4. If you want mechanical breaths to be triggered in the
phase during which no chest compression is carried out:
Increase the trigger level and/or reduce PEEP.
or
If the trigger is not activated or activated too infrequently and
no compressions are detected: Reduce the trigger level and/or
increase PEEP.
5. If the patient experiences spontaneous circulation again:
Stop chest compressions.
or
Cancel the CPR CCSV mode with the CPR button .
Result Ventilation is carried out in CPR CCSV mode.